This report describes a complete multimodal image registration pipeline for four medical images:
MRI_T2.dcmMRI_DWI.dcmPET.dcmMRI_T2_rot.dcm
The workflow includes image loading, spatial-resolution harmonization, zero padding, exhaustive translation search, exhaustive rotation search, visual assessment with checkerboard overlays, and evaluation through SSD, NCC, and Mutual Information (MI).
The goal is to align multiple imaging modalities into a common spatial frame so that anatomical structures can be compared consistently across acquisitions.
Main tasks:
- Load the DICOM images and inspect size and resolution.
- Resample floating images to match reference spatial resolution.
- Apply zero padding to obtain common image size.
- Estimate best translation for DWI and PET relative to T2.
- Estimate best rotation for
MRI_T2_rotrelative toMRI_T2. - Evaluate results using SSD, NCC, and MI.
- Visualize final alignment with checkerboard overlays and summary figures.
The T2-weighted MRI image is used as the reference image because:
- it has the highest in-plane resolution,
- it provides detailed anatomy for geometric matching.
| Image | Size (pixels) |
|---|---|
| MRI T2 | 512 x 512 |
| MRI DWI | 128 x 128 |
| PET | 128 x 128 |
| MRI T2 Rot | 512 x 512 |
| Image | Pixel spacing (mm) |
|---|---|
| MRI T2 | 0.449 x 0.449 |
| MRI DWI | 1.750 x 1.750 |
| PET | 2.5743 x 2.5743 |
| MRI T2 Rot | 0.449 x 0.449 |
Floating images were resampled to match T2 spacing.
- DWI -> T2:
3.8976 - PET -> T2:
5.7333
| Image | Size after scaling |
|---|---|
| MRI T2 | 512 x 512 |
| MRI DWI (scaled) | 499 x 499 |
| PET (scaled) | 734 x 734 |
After scaling, images were zero-padded to a common field of view.
Translation registration was performed for:
- T2 vs DWI
- T2 vs PET
Search range:
dx, dy in [-20, 20]
Metrics computed at each shift:
- SSD
- NCC
- MI
Final translation was selected from NCC optimum.
- minimum SSD:
1947122952.0000 - maximum NCC:
0.9329 - maximum MI:
0.9518
- minimum SSD:
2192681244.0000 - maximum NCC:
0.9649 - maximum MI:
0.8851
| Floating image | Best translation [x y] (pixels) |
|---|---|
| MRI DWI | [-1, -1] |
| PET | [6, -16] |
Rotation registration was performed between:
MRI_T2.dcmMRI_T2_rot.dcm
Angular search:
- range:
0 degto359.5 deg - step:
0.5 deg
All three metrics identified the same optimal angle:
- SSD minimum at
340.00 deg - NCC maximum at
340.00 deg - MI maximum at
340.00 deg
| Registered pair | Final MI |
|---|---|
| T2 vs T2_rot_registered | 2.9144 |
| T2 vs DWI_registered | 0.9471 |
| T2 vs PET_registered | 0.8837 |
| Image | Scaling factor | Translation [x y] (pixels) | Rotation |
|---|---|---|---|
MR_T2.dcm |
REF | [0, 0] |
- |
MRI_DWI.dcm |
3.8976 | [-1, -1] |
- |
PET.dcm |
5.7333 | [6, -16] |
- |
MRI_T2_rot.dcm |
- |
- |
340.00 deg |
Medical_Image_Registration.m- full MATLAB script for the registration workflow.lib/metrics/SSD.m- Sum of Squared Differences.lib/metrics/NCC.m- Normalized Cross Correlation.lib/metrics/MI.m- Mutual Information.lib/operators/zeroPadding.m- field-of-view harmonization helper.lib/visualization/checkerboard_view.m- visual comparison helper.figures/- full set of exported registration figures.
The workflow successfully performs multimodal registration between T2, DWI, PET, and rotated T2 images.
Key outcomes:
- T2 was used as the anatomical reference.
- DWI and PET were resampled and padded before translation search.
- Best translations were
[-1, -1]for DWI and[6, -16]for PET. - Best rotation for
MRI_T2_rotwas340.00 deg. - Heatmaps, checkerboards, and summary reports confirmed good final alignment.